8-6
8.4.2.3 Setting Arrh. Threshold
The threshold of PVCs high limit, tachy and others can be set in “Arrh. Threshold Setup”
sheet.
8.4.2.4 Others
In “Others” tab sheet, the alarm latch can be set here. Please refer to 8.9 Latching Alarms for
details.
8.5 Pausing Alarms
For the monitors supporting bidirectional control of alarm pause, you can click the Alarm
Pause icon
in the viewbed screen or Alarm Pause in the drop-down menu in the
multibed screen to pause all alarms of current monitor for a period of time, which can be set
on the monitor. After alarm pause time ends, the system will automatically release the alarm
pause. You can also release the alarm pause by clicking the icon
or Alarm Pause in the
drop-down menu again.
You can enable or disable the remote alarm control function on the CMS. When remote alarm
control is disabled, the icon
is inactive and the “Alarm Pause” control is not available in
the drop-down menu in the multibed screen. Refer to Section 12.2.8 Other to see how to
enable or disable the remote alarm control function.
8.6 Silencing Alarms
For the monitors supporting bidirectional control of alarm silence, you can enter the status of
Silence through following ways:
In the CMS viewbed screen, click the Silence icon
.
In the multibed screen, you can click the right button of mouse when the cursor is in the
window where the patient needs the setting of Silence, or select the option ‘Silence’ in
its drop-down menu.
When Silence is selected, both the monitor and corresponding patient window on the CMS
will enter the alarm silenced status: after the physiological alarm is silenced, appears
before the alarm message; for some technical alarms, appears before the alarm message
after the alarm is silenced; for some technical alarms, the alarm message changes to prompt
message after the alarm is silenced; for some other technical alarms, the alarm message is
cleared after the alarm is silenced.