8-1
8 Alarm Control
8.1 Alarm Structure
Alarms, triggered by a vital sign that appears abnormal or by technical problems of the
monitor, are sent to the CMS by the monitors and then indicated to the users by the CMS.
Most of the alarms originally come from the monitors. The alarms coming from the CMS
itself are displayed in the patient window’s technical alarm area or system alarm area on the
upper screen. Please refer to Appendix B Central Monitoring System Alarms for the alarm
messages.
By nature, the alarms are classified into two categories:
Physiological alarms, and
Technical alarms.
By severity, the alarms are classified into:
High priority alarms,
Medium priority alarms, and
Low priority alarms
NOTE
z For details on alarms, refer to the monitor’s operation manual.
8.2 Alarm Mode
The CMS gives audible and visual alarms in compliance with international standards.
WARNING
z The audible and visual alarms given by the CMS comply with the 60601-1-8
standard. The hospital or institution employing the use of the CMS should give
adequate training to the operators.