7-4
7.2.3 Parameter Area
This area is to display parameter values transmitted from the monitor. Each parameter
module occupies a rectangular area, which is hereafter called parameter window. In the
single-screen mode, this area can only accommodate part parameter values due to space
confines of the screen. However, you can drag the vertical scroll bar to view more.
Click in a parameter window. The border of the window will highlight when the window is
selected. Clicking on it again will enter the “Parameter Setup” tab sheet. For details, refer to
Part 7.3 Parameter Setup.
NOTE
z The waveforms and parameters are consistent between the CMS and the
monitor.
z If you change the display order of waveforms and parameters on the monitor,
the display order of waveforms and parameters on the CMS will not change
automatically. Vice versa.
z In case a module of the monitor is turned off or a parameter is not being
monitored, its corresponding waveforms and parameters displayed on the
CMS’s screen will disappear at the same time.
7.3 Parameter Setup
Click in a parameter window, and the border of this window will appear highlighted. Clicking
again will open the “Parameter Setup” window. The current tab sheet depends on the
parameter window you have selected. For example, if you click in the ECG parameter
window, the Parameter Setup window will open, with the “ECG” tab sheet at the front.
In this tab sheet, you can set alarm, record switch and sweep speed for all parameters of ECG,
SpO
2
, C.O., etc. For details on alarms, refer to Part 8.4 Alarm Setup.