12-10
12.2.7 Telemetry
If your CMS is connected with telemetry monitors, select the “Telemetry” tab, you can set the
following items in an opened tab sheet.
System Information
In the “System Information” area, the mainboard software version, receiveboard software
version, total run-time and “IP Address” are displayed. Among them, the “IP Address” can be
modified.
Alarm Level Setup
Click the “Alarm Level Setup” button. In the popup menu, you can set the “Alm Lev of ECG
Lead Off” and “Alm Lev of SpO2 Sensor Off” to “High”, “Med” or “Low”. The default
setting is “Low”.
Monitoring Mode Setup
Monitoring mode setup means to establish corresponding relations between the telemetry
transmitter and multibed patient sector. The CMS provides two monitoring modes:
Fixed monitoring: to establish fixed corresponding relations between the telemetry
transmitter and multibed patient sector. Such corresponding relations are not changed when
the patient is discharged. The patient sector displays bed number and serial number of
telemetry transmitter and the bed mark color is not changed. The patient sector displays “No
Patient Admitted”.
Flexible monitoring: no fixed corresponding relations are established between the telemetry
transmitter and multibed patient sector. The corresponding relations are relieved when the
patient is discharged. The patient sector is idle and displays “Idle sector. Click to admit a
patient”.
Click the “Monitoring Mode Setup” button and select the telemetry monitoring mode from
the pop-up menu. The options are “Fixed Monitoring” and “Flexible Monitoring”. The setting
is activated immediately.