12-11
Frequency Setup
Select frequency range:
1. You can select the appropriate frequency segment from “Freq Segment”. The minimum
“Start Freq” and maximum “End Freq.” corresponding to the frequency segment will be
displayed automatically.
2. You can also click the drop-down menu to select other “Start Freq” and “End Freq.”.
Frequency setup list:
This area shows channel number, channel name, and the transmitter frequency and signal
strength corresponding to each receiver channel in the tabular form.
1. Select a channel from the list. If a frequency point is currently available and it is not
used by other channel, you can click the “Program…” button to program this channel.
2. If the serial port connects no corresponding transmitter at this time, the message will be
displayed.
3. If the serial port is properly connected with the corresponding transmitter, the Program
screen will be displayed.
The default value of “SN” is the serial number that has been written in the transmitter.
The default value of the “Frequency” is the transmitter frequency corresponding to the
current receiver channel. You can either select other frequency by clicking the
drop-menu button.
The default value of “Power” is 23. You can either select other power value by clicking
the drop-menu button.
4. Select “OK”, and you will modify the configuration of this transmitter.
5. After the configuration is completed, click “OK” in the following dialog box.
Discharge a patient
Select a channel from the frequency setup list. If this channel has admitted a patient, you can
click the “Discharge” button to discharge the patient.
Settings of Lethal Arrh. OFF
The “Lethal Arrh. OFF” can be set for telemetry monitoring system;
If it is set to “Disable”, the switches of all lethal Arrhythmia alarms are set to “ON” and
unchangeable. “All Alarm Off” button is inactive.
If it is set to “Enable”, the switches of lethal Arrhythmia alarms are able to be set. “All
Alarm Off” button is active.