C-11
exceeds 1 second.
AG SELF-TEST ERR Yes No 1 AG module fails or
fails to communicate
with the host.
Unplug and re-plug the
module and then restart
the monitor, or plug the
AG module in another
monitor to check if it
operates properly.
AG HARDWARE MALF No No 1 AG module hardware
failure. The module
enters the standby
mode or stop AG
monitoring.
Remove the AG module
from the monitor and
send for authorized
maintenance personnel.
AG HARDWARE ERROR Yes No 2 AG module hardware
fault. The module
remains AG
monitoring.
AG DATA LIMIT ERROR Yes No 2 AG module fault.
AG ZREF FAIL Yes No 3 Failure in AG module
zeroing.
AG CAL FAIL Yes No 1 Failure in AG module
calibration.
Unplug and re-plug the
module and then restart
the monitor.
CO2 ACCURACY
UNSEPCIFIED
Yes No 3
N2O ACCURACY
UNSEPCIFIED
Yes No 3
O2 ACCURACY
UNSEPCIFIED
Yes No 3
HAL ACCURACY
UNSEPCIFIED
Yes No 3
ENF ACCURACY
UNSEPCIFIED
Yes No 1
ISO ACCURACY
UNSEPCIFIED
Yes No 3
SEV ACCURACY
UNSEPCIFIED
Yes No 3
DES ACCURACY
UNSEPCIFIED
Yes No 3
AwRR ACCURACY
UNSEPCIFIED
Yes No 3
The measurement is
out of specified range
at specified accuracy.
Check that appropriate
gas is used and the
clinical environment is
stable. If the problem
persists, send for
authorized maintenance
personnel.