Mindray DPM 1 Manual del operador Pagina 128

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 274
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 127
10-24
10.6.3 Arrhythmia Analysis Menu
Select ARR ANALYSIS >> in ECG SETUP menu. The following menu appears. The menu items
vary with the algorithm packages.
Figure 10-16 Arrhythmia Analysis (Mortara algorithm package)
In this menu, you can perform the following settings:
ARR ANAL Arrhythmia analysis
ON: Enables the arrhythmia analysis;
OFF: Disables the arrhythmia analysis.
PVCs ALM PVCs alarm
ON: When the PVCs alarm occurs, the monitor gives alarm
indications and saves the alarm;
OFF: When the PVCs alarm occurs, the monitor neither gives
alarm indications nor saves the alarm;
When OFF is selected, the
icon is displayed on the right of
PVCs in the parameter window.
ALM LEV Alarm level
Options: HIGH, MED and LOW.
ALM REC Alarm recording
ON: The monitor starts recording when the PVCs alarm occurs.
OFF: The monitor does not record when the PVCs alarm
occurs.
ALM HI Upper alarm limit
Determines the upper limit of the PVCs alarm. Range: 110.
An alarm is triggered when the PVCs exceeds the upper limit.
Vista de pagina 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 273 274

Comentarios a estos manuales

Sin comentarios