A-2
A.3 Power Source Specifications
AC Power Supply Specifications
Input voltage 100 to 240 V~
Current 1.4A to 0.6A
Frequency 50/60 Hz
Fuse T 3 A, 250 V
Internal battery
Number of batteries 2
Type Sealed lead-acid battery or lithium-ion battery
Time to shutdown >5 min (after the first low-power alarm)
Sealed lead-acid battery
Nominal voltage 12 VDC
Capacity 2.3 Ah
Operating time
48 minutes or 120 minutes typical when powered by one or two
new fully-charged batteries respectively (25, ECG, SpO2,
NIBP measurement per 15 minutes).
Charge time
A maximum of 6 h for each battery, and a maximum of 12h for
both (in the running status or standby mode)
Lithium battery
Rated voltage 11.1 VDC
Capacity 4.4 Ah
Operating time
120 minutes or 300 minutes typical when powered by one or two
new fully-charged batteries respectively (25, ECG, SpO2,
NIBP measurement per 15 minutes).
Charge time A maximum of 6.5h (in the running status or standby mode)