Mindray DPM 1 Manual del operador Pagina 150

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 274
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 149
11-8
SpO
2
and PR alarm limits:
Parameter Upper Limit Lower Limit Step
SpO2 (lower limit+1) ~100 0~(upper limit-1) 1
PR (lower limit+2) ~254 0~(upper limit-2) 1
The SpO
2
and PR alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Adult 100 90
Pediatric 100 90
SpO2
Neonate 95 90
Adult 120 50
Pediatric 160 75
PR
Neonate 200 100
WARNING
z Setting the SpO2 upper alarm limit to 100% will disable the upper alarm limit.
High oxygen levels may predispose a premature infant to retrolental fibroplasia.
Therefore, the upper alarm limit for oxygen saturation must be carefully selected
in accordance with the commonly accepted clinical practices.
SWEEP Waveform speed
Options: 12.5 and 25.0 mm/s.
PR SOUND Pulse volume
Range: 010. 0 indicates the volume is closed and 10
indicates the maximum volume.
SENSITIVE Sensitivity of SpO2 calculation
Options: HIGH, MED and LOW.
When HIGH is selected, the monitor gives quick response to
the change of the oxygen saturation but the measurement
accuracy might be compromised.
When LOW is selected, the monitor gives slow response to
the change of the oxygen saturation but the accuracy is
relatively high.
NIBP SIMUL ON: When measuring NIBP and SpO2 at the same side, the
SpO2 alarm status remains unchanged until the NIBP
measurement is finished;
OFF: The function of NIBP SIMUL is disabled.
Vista de pagina 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 273 274

Comentarios a estos manuales

Sin comentarios