C.2.4 TEMP Module Alarm Messages
Alarm message A B L Cause Measure
TEMP1
SENSOR OFF
No Yes 3 The T1 sensor is not
connected correctly to the
patient or the monitor.
Check for correct
connection of the T1
sensor.
TEMP2
SENSOR OFF
No Yes 3 The T2 sensor is not
connected correctly to the
patient or the monitor.
Check for correct
connection of the T2
sensor.
TEMP
SELFTEST
ERROR
No No 1 Circuit fault of the
temperature channel.
Contact our company for
repair.
TEMP
CALIBRATION
ERR
No No 2 Error in temperature channel
calibration.
Restart the monitor. If the
error remains, contact our
company for repair.
C.2.5 NIBP Module Alarm Messages
Alarm message A B L Cause Measure
NIBP
SELFTEST ERR
Yes Yes 1 Error in NIBP
initialization.
Select RESET from the
NIBP SETUP menu. If the
error remains, contact our
company for repair.
LOOSE CUFF No Yes 3 The NIBP cuff is not
properly connected.
AIR LEAK No Yes 3
PNEUMATIC
LEAK
No Yes 3
Leak in the airway.
CUFF TYPE
ERR
No Yes 2 The cuff applied is not
appropriate to the patient
type.
AIR PRESSURE
ERROR
No Yes 3
WEAK SIGNAL No Yes 3
SIGNAL
SATUATED
No Yes 3
Check the patient’s
condition, and check if the
patient type is correct.
Replace with a proper cuff
and connect it correctly. If
the problem still exists,
contact our company for
repair.
RANGE
EXCEEDED
No Yes 3
Failure occurs in the pulse
measurement. The monitor
cannot perform
measurement, analysis, or
calculation.
EXCESSIVE
MOTION
No Yes 3 Excessive motion of the
patient’s arms.
OVER
PRESSURE
No Yes 2 The airway might be
blocked.
Check the patient’s
condition, and check if the
patient type is correct.
Replace with a proper cuff
and connect it correctly. If
the problem still exists,
contact our company for