Mindray DPM 1 Manual de servicio Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
Circuit Principle
Pulse oximeter Service Manual 2-8
2.3 EMC Design
The following protections are used for EMC design in DPM1:
1. Every IC has one by-pass capacitor, some ICs have different rating capacitors to
enlarge the filter band;
2. Serial resistors on the communication lines connected with the PC, to reduce the
interference propagation;
3. RC low pass filter on most low frequency signals, to reduce high frequency
interference;
4. One ground plane, no discrimination between digital ground and analog ground;
5. Match resistor used on important signals, e.g. read/write, chip select etc., to
prevent oscillating or overshoot
Vista de pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 33 34

Comentarios a estos manuales

Sin comentarios