Circuit Principle
2-1 Pulse oximeter Service Manual
Chapter 2 Circuit Principle
2.1 Overview
The DPM1 pulse oximeter collects SpO2 data from the sensor and sends to mainboard.
The maniboard processes the data and displays the results (SpO2 values) on the LCD
screen. Pulse-strength bar, battery remained capacity and data export indication are also
shown on the screen. DPM1 can be connected with PC through serial port for data
transportation and the data can be printed out from PC.
2.1.1 Hardware Theory
DPM1’s mainboard consists of power circuit, main logic circuit, display circuit and control
input circuit. The SpO2 value can be displayed on the LCD or be exported to PC through
serial port. The data also can be saved in the EEPROM on the main board as history
record. Watchdog circuit is used to reduce interference. Low-power design is adapted to
the main board in order to saving energy.
s1c133209
Power Monitor circuit
Keyboard Interface
LCD
PC Comunication port
Debug port
Switch Circuit
Infrade drive circuit
WDT data Memory
Flash
A/D Convert
Fig 2-1 DPM1 Pulse oximeter Block Diagram
2.1.1.1 CPU Power System
S1C33209 uses two voltages. VDD is used to supply the CPU and VDDE is the power