Circuit Principle
Pulse oximeter Service Manual 2-4
2.5V voltage shift should be required. R27-C20, R20-C30 and R23-C19 form RC low pass
filters to reduce the interference from high frequency noise. C31 is used to prevent
self-oscillating and enhance the reliability.
C19
1000p
R23 1K 1%
AGND
SIGNAL
R20 10K 1%
R11
100K 1%
C30
10P
C31 10P
R12 100K 1%
R26
9.09K 1%
R27 9.09K 1%
C20
1000p
AGND
OFFSET
5
6
7
U4B
LMC6482
T51
tp
VREF2.5
VIN
Fig 2-5 Voltage Shift Circuit
Due to the input signal limits of 0V~5V for the ADC chip, a clamp diode is used at the
SIGNAL input circuit.
2.1.1.6 LCD Display Module
DPM1 uses LCD module to display data and patient information. The interface between
LCD and CPU includes power line, data bus, read/write control, chip selection and
address line.
Backlight of the LCD is separated from others, which is a LED and controlled by a
CPU-controlled transistor or MOS transistor.
A π-type filter was adopted for this power in order to reduce the interference from the
voltage multiplier circuit in the LCD module.
2.1.1.7 Switch ON/OFF Delay Circuit
DPM1 has no delay for switch on, but there is 2s delay for switch off.
In the delay circuit, a voltage of +7V is slowly supply to the Q5. The RC network on Q4’s
base is used to eliminate the button-press shaking and to protect the circuit from
electrostatic charge. Before power on, Q5 is off and does not influence Q4. When the Vbe
of Q5 goes up and over 0.7V, the Q5 will work. Additionally, the capacitance of E104
should be large enough, otherwise the time interval between 0V and 0.7V will be short.
When power off, the discharge time of the 10uF capacitor will become short because the
diode D7 is used to fasten the charge release. After power on, Q5 is on and the base of
Q4 is 0V. When a button is pressed, Q4 is off and RST-POWER will not function. D2 and