Mindray TRIO Manual de servicio Pagina 31

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 82
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 30
Calibration and System Checks Calibration/Maintenance
2 - 6 0070-10-0591-01 Trio™ Service Manual
NIBP CAL
Connect a test chamber (P/N 0138-00-0001-03) and a calibrated digital mercury
manometer with bulb via T fitting to the NIBP quick connect fitting of the Trio Monitor
(Figure 2-5.)
1. From the MAINTENANCE menu select NIBP CAL. The selected tile will now read
STOP CAL. The user may select this tile at any time to stop the test.
2. Using the bulb, inflate the pneumatic system so that the digital mercury manometer reads
0, 50 and 200 mmHg in turn. The difference between the indicated pressure of the
digital mercury manometer and the indicated pressure in the NIBP parameter tile of the
Trio Monitor should not exceed 3 mmHg. If it exceeds 3 mmHg, please contact a
service technician.
FIGURE 2-5 Diagram of NIBP Calibration
DIGITAL OR
MERCURY
MANOMETER
500 ml.
TEST CHAMBER
P/N 0138-00-0001-03
BULB
hose
TRIO MONITOR
NIBP
Vista de pagina 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 81 82

Comentarios a estos manuales

Sin comentarios