Trio™ Operating Instructions 0070-10-0666-01 2 - 17
Operations Parameter Menus
Standard 5-wire Lead Sets
Standard 5-wire lead sets monitor 7 ECG leads: I, II, III, aVR, aVL, aVF and V.
FIGURE 2-16 5-wire Lead Placement
(AHA)
FIGURE 2-17 5-wire Lead Placement
(IEC)
Place the RA (white) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
Place the LA (black) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
Place the LL (red) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Place the RL (green) electrode on the
patients lower right abdomen within the
rib cage frame.
Place the V (brown) electrode in one of
the V-lead positions (V1 – V6) depicted
in the following section.
Place the R (red) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
Place the L (yellow) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
Place the F (green) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Place the N (black) electrode on the
patient’s lower right abdomen within the
rib cage frame.
Place the C (white) electrode in one of
the C-lead (C1 – C6) positions depicted
in the following section.
Black
White
RA
RL
Brown
V Lead
(any V position)
LA
Green
LL
Red
V
Yellow
Red
R
N
White
C Lead
(any position)C
L
Black
F
Green
C