12. Battery indicator
It indicates the status of the battery. For details, refer to 2.4 Battery.
13. AC power indicator
ON: It indicates that the AC power is applied to the monitor;
OFF: It indicates that the monitor is not applied to the monitor.
14. Patient type indicator
It indicates the patient types: respectively adult, pediatric and neonate from left to
right.
15. Temperature site and mode indicator
It indicates the temperature measurement site and monitoring mode: respectively oral,
axillary, rectal and monitor.
16. Pulse strength indicator
It indicates the pulse strength of a patient.
17. RECORD
Press this key to start or stop the recording.
18. NIBP status indicator
ON: It indicates that the monitor is performing the NIBP measurement;
OFF: It indicates that the monitor is not performing the NIBP measurement.
19. NIBP
Press this key to start an NIBP measurement, or press this key during measurement to
stop it.
20. SILENCE
Shortly pressing this key pauses the current alarm for 2 minutes. Pressing and holding
this key for more than 2 seconds disables audible alarms, thus entering the alarm
silenced mode. During alarm silenced, the system will return to the normal status when
a new alarm occurs.
21. Silence indicator
OFF: normal status; in this status, when an alarm occurs, the system can give an
audible alarm according to the alarm level;
ON: alarm silenced status; in this status, the system cannot give audible alarm.
FLASH: alarm paused status; in this status, the system cannot give the audible
and visual alarm.
22. Up
Press this key to move the cursor upward.